InfiniCon Systems VAP016 User Manual Page 2

  • Download
  • Add to my manuals
  • Print
  • Page
    / 4
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 1
Installation
Vacuum Connection
Skilled personnel
The vacuum connection may only be established
by persons who have suitable technical training
and the necessary experience or who have been
instructed by the end-user of the product.
DANGER
Caution: overpressure in the vacuum system
>1 bar
Injury caused by released parts and harm
caused by escaping process gases can result if
clamps are opened while the vacuum system is
pressurized.
Do not open any clamps while the vacuum
system is pressurized. Use the type clamps
which are suited to overpressure.
Caution
Caution: dirt sensitive area
Touching the product or parts thereof with one's
bare hands increases the desorption rate.
Always wear clean, lint-free gloves and use
clean tools when working in this area.
Caution
Caution: vacuum component
Dirt and damages impair the function of the
vacuum component.
When handling vacuum components, take
appropriate measures to ensure cleanliness and
prevent damages.
Caution
Keep the protective lids and put them in place
again when removing the product from the
vacuum system.
Remove the protective lids and install the valve to the
vacuum system by means of the small flange fittings.
Any installation angle and flow direction may be chosen.
Seal with
centering ring
Clamp
ISO-KF
connection flange
Protective lid
Compressed Air Connection
Skilled personnel
The compressed air connection may only be
established by persons who have suitable tech-
nical training and the necessary experience or
who have been instructed by the end-user of the
product.
Caution
Specifications for the plastic tube:
· ø4 mm, ø6 mm or ø¼“
· bursting pressure ³ 10 bar overpressure
(1 MPa)
· material: PA soft or PU.
Caution
The compressed air must meet the following
specifications:
· free of particles >5 µm
· DN 16 and DN 25: 3 … 5 bar overpressure
DN 40: 3 … 6 bar overpressure
· dry, free of oil or containing oil
(keep using the same quality).
If compressed air containing oil is used, dispose
of the exhaust compressed air outlet in
accordance with the relevant regulations.
Caution
To ensure leak tightness of the instant push-in
fitting
· cut the plastic tube square
· make sure the outside of the plastic tube is
not damaged.
Caution
Caution: plastic thread
The plastic thread is damaged by tilting or
overturning the instant push-in fitting.
· use the enclosed instant push-in fitting (with
extra-long thread) only.
· screw in the instant push-in fitting without
tilting it and without exceeding the tightening
torque of 0.5 Nm.
Compressed Air Inlet
Caution
Contrary to the illustration below, the compres-
sed air inlets and outlets of the following angle
valves are reversed in position:
Vacuum connection DN 40 ISO-KF and serial
number (F-No) from
· 101 for valve with aluminum housing
· 118 for valve with stainless steel housing
If you wish to connect a ø6 mm, or ø¼“ plastic tube,
exchange the instant push-in fitting.
Tightening torque 0.5 Nm
Compressed air
inlet
Compressed air
outlet
Insert the tube into the instant push-in fitting until the
mechanical stop is reached. Check that it is correctly
mounted by slightly pulling.
Compressed air
inlet
Compressed air
outlet
Plastic tube
Compressed Air Outlet
Caution
Contrary to the illustration below, the compres-
sed air inlets and outlets of the following angle
valves are reversed in position:
Vacuum connection DN 40 ISO-KF and serial
number (F-No) from
· 101 for valve with aluminum housing
· 118 for valve with stainless steel housing
Screw in the enclosed instant push-in fitting for exhausting
the compressed air if necessary. Push the tube into the
instant push-in fitting until the mechanical stop is reached.
Check that it is correctly mounted by slightly pulling.
Tightening
torque 0.5 Nm
Compressed air outlet
Compressed air inlet
Electrical Connection
Skilled personnel
The electrical connection, in accordance with the
VDE 0100 guidelines, may be made only by a
licensed electrician, qualified as per VDE 0105.
The line cables shall be isolated from the line
supply during all electrical work.
WARNING
Caution: mains voltage
The pilot valve can get destroyed if a wrong
mains voltage is applied.
The local mains power rating must correspond
with the nominal voltage of the pilot valve (see
product nameplate). If they do not correspond,
exchange the pilot valve (® Furter information).
Caution
The cable must meet the following specifications:
· flexible
· conductor cross-section £0.75 mm
2
· cable diameter £10 mm
· 6-pole without protective conductor or
7-pole with protective conductor.
Preparing the connector
Slide the screw fitting, connector housing, and strain
relief on the cable.
Screw fitting
Connector housing
Strain relief
Seal
Washer
Hex head screw
si ma60e1- b
(0210)
Original: German sima60d1-b (0210)
Page view 1
1 2 3 4

Comments to this Manuals

No comments