Angle valve
pneumatically actuated
bellows sealed
with position indicator and pilot valve
normally open
VAP016 … 040-A/X
Instruction Sheet
Incl. Manufacturer's Declaration
sima60e1-b (0210)
Product Identification
In all communications with INFICON, please specify the infor-
mation on the product nameplate. For convenient reference
copy that information into the space provided below.
INFICON, FL - 9496 Balzers
Typ:
No:
F-No:
Nominal voltage
Validity
This document applies to products with the following part
numbers:
Aluminum housing:
DN 16
ISO-KF
DN 25
ISO-KF
DN 40
ISO-KF
Nominal voltage
(pilot valve) N.O.
250-206 250-226 250-246
24 VDC (=)
Stainless steel housing:
DN 16
ISO-KF
DN 25
ISO-KF
DN 40
ISO-KF
Nominal voltage
(pilot valve) N.O.
250-216 250-236 250-256
24 VDC (=)
The part number can be taken from the product nameplate.
If not indicated otherwise in the legends, the illustrations in
this document correspond to the valve with the nominal
diameter DN 25 ISO-KF. They apply to valves with other
nominal diameters by analogy.
We reserve the right to make technical changes without prior
notice.
Intended Use
The angle valves are used as shut-off and venting devices for
vacuum applications.
Description
Position indication
(visual)
Position indicator
(electrical)
Connector
(enclosed)
Vacuum connection
DN 16 - 40 ISO-KF
Control valve
Pneumatic
actuator
Instant push-in fittings
(enclosed)
2x
2x
Functional principle
When the pilot valve is activated, the angle valve is closed by
the pressure spring. The position indicator is invisible.
When the pilot valve is deactivated, the angle valve is opened
by the pneumatic actuator. The green position indicator be-
comes visible.
The final positions can be polled by the electrical position
indicator.
Safety
Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and
environmental damage.
Caution
Information on correct handling or use. Disregard can lead
to malfunctions or minor equipment damage.
20
Dimensions in mm
General Safety Instructions
· Adhere to the applicable regulations and take the neces-
sary precautions for the process media used.
Consider possible reactions between the materials and the
process media.
· Adhere to the applicable regulations and take the neces-
sary precautions for all work you are going to do and
consider the safety instructions in this document.
· Before beginning to work, find out whether any vacuum
components are contaminated. Adhere to the relevant
regulations and take the necessary precautions when
handling contaminated parts.
Communicate the safety instructions to all other users.
Liability and Warranty
INFICON assumes no liability and the warranty becomes null
and void if end-user or third parties
· disregard the information in this document
· use the product in a non-conforming manner
· make any kind of interventions (modifications, alterations
etc.) on the product
· use the product with accessories and options not listed in
the corresponding product documentation.
The end-user assumes the responsibility in conjunction with
the process media used.
Technical Data
Pilot valve
connection
type
nominal voltage
power
duty cycle
nominal diameter
soldered joints
normally open
see product nameplate
1 W
100%
0.42 mm
Position indicator
connection
rating
soldered joints
250 VAC / 25 VA / 0.1 A
50 VDC / 12.5 W / 0.25 A
Connection flange DN 16
ISO-KF
DN 25
ISO-KF
DN 40
ISO-KF
Actuation opening: pneumatic
closing: by pressure spring
Compressed air supply
tube connection ø4 mm, ø6 mm or ø¼“
pressure range
(overpressure)
3 … 5 bar
3 … 6 bar
piston displacement 5.5 cm
3
12.1 cm
3
26.2 cm
3
Stroke of the valve plate 5 mm 10 mm 14 mm
Conductance
1)
4.5 l/s 16 l/s 40 l/s
Switching frequency
2)
Opening time
2)
Closing time
2)
100 / min
100 ms
200 ms
100 / min
110 ms
290 ms
75 / min
150 ms
250 ms
Cycle life
3)
10 million
Tightness 1×10
-9
mbar l/s
Pressure range min. 1×10
-8
mbar
Pressure range max.(abs.) 4 bar 2.5 bar
Pressure difference Dp
in closing direction
in opening direction
Opens to a pressure
difference Dp
4)
4 bar
2 bar
4 bar
2 bar
1.5 bar
2 bar
Temperatures
ambiance
bakeout
housing
aluminum
stainless steel
actuator / pilot valve
0 °C … 50 °C
80 °C
150 °C
50 °C
Type of protection
Protection class
IP 50 according to DIN 40 050
II
Installation angle
Flow direction
any
any
Materials
housing
aluminum
stainless steel
bellows / valve plate
pressure spring
DN 16 + 25 ISO-KF
DN 40 ISO-KF
seals
shell / cylinder unit
protective lids
packing material
3.2572
1.4301
1.4541 / 1.4301
1.4301
1.1200
FPM
PBTP
PE
carton box, PE, PU
Weight
housing
aluminum
stainless steel
0.3 kg
0.4 kg
0.44 kg
0.75 kg
0.9 kg
1.6 kg
1)
For air with molecular flow
2)
With pressure difference Dp=0 and compressed air = 5 bar
(overpressure)
3)
Cycles without expendable parts (seals) and under clean
operating conditions
4)
Compressed air = 5 bar (overpressure)
DN
DN
DN 16 ISO-KF
DN 25 ISO-KF
DN 40 ISO-KF
A
154
176
196.5
B
127
147.4
167.3
C
71
92.5
112
D
60
74
98
E
51
63
83
DN F G
100
108
120
40
50
65
M5
Dimensions
2x
2x
Page view 0
1 2 3 4

Summary of Contents

Page 1 - Angle valve

Angle valvepneumatically actuatedbellows sealedwith position indicator and pilot valvenormally openVAP016 … 040-A/XInstruction SheetIncl. Manufacturer

Page 2 - Installation

InstallationVacuum ConnectionSkilled personnelThe vacuum connection may only be establishedby persons who have suitable technical trainingand the nece

Page 3 - Vacuum Connection

 Skin the cable and mount the insulating sleeves ifrequired.Insulating sleeves(not incuded) Solder the cable. Slide the insulating sleeve over theso

Page 4

CautionCaution: dirt sensitive areaTouching the product or parts thereof with one'sbare hands increases the desorption rate.Always wear clean, li

Comments to this Manuals

No comments